Mv6s8tal.gif (32757 bytes)

ED-520XL
Flaw Detector

 

 

  • Highly sensitive resonance-type eddy current instrument
  • Proven, reliable microprocessor-based circuit design
  • Wide selection of probes and coils - interchangeable with existing equipment
  • Includes test block and pencil probe
  • Variable frequency range from 55 KHz to 220 KHz
  • User-selectable 115 or 230 VAC operation
  • Dust-proof, weather resistant rugged aluminum enclosure for operation under most environmental conditions
  • Extremely versatile, yet simplified operation


The ED-520XL is a microprocessor-based electronic Eddy Current instrument which offers high sensitivity, versatility and simplicity of operation - all in a self-contained unit.

The instrument will locate surface and near-surface defects in non-magnetic materials and surface flaws in magnetic materials. The ED-520XL will sort magnetic and non-magnetic materials according to hardness, alloy type, carbon content, heat treat condition and tensile strength. It will also measure coating thickness and conductivity on a comparison basis.

Whether you are checking for cracks in aluminum, or, determining the condition of heat-treated parts, the ED-520XL accurately measures the status of your test part. This Eddy Current instrument is widely used to detect flaws in aircraft and automobile parts like wheels, struts, engine components and engine mounts. In foundries and manufacturing, the ED-520XL is used for quality testing of in-coming raw materials and finished product quality control. The ED-520XL is also used in research and maintenance to verify test results and check equipment for fatigue.


The ED-520XL is easy to operate with its push-button controls. In addition to the easy-to-read meter, you can set an alarm to notify you when the meter hits a given level. Or, you can attach an external accessory for additional flexibility. For complex testing applications, we have designed and produced custom probes for automotive and aviation testing, such as the wheel bead seat and bolt hole inspection and many manufacturing applications. The control dials for Sensitivity, Lift-Off/Frequency, and Balance contribute to your testing sensitivity level and testing calibration.

The instrument operates in a variable frequency range of 55 KHz to 220 KHz to accommodate a wide range of testing materials.

The instrument comes with an aluminum flaw sensitivity test block on the face of the unit. Also included in the package is a medium frequency, short pencil probe and cable.

The front panel complement includes an easily read meter and the following controls and components:

Sensitivity: Single-turn control for continuous adjustment of the gain between push-button selectable settings.
Charge: LED indicates line cord is plugged in and battery is being charged.
Frequency: Continually varies adjustment of operating frequency over the range of 55 KHz to 220 KHz. In the automatic set-up mode, used to fine adjust lift-off/frequency.
Balance: Controls the position of the meter needle.
Gate: A rotary switch to select upscale or downscale triggering for the threshold gate The single-turn potentiometer positions where the threshold gate triggers. The LED indicator lights when the gate has been triggered.
Accessory: Analog outputs available for chart recorder; TTL output for relay.
Probe: BNC-type connector for a wide range of probes and coils. All previous ED-520 and ED-530 probes and coils are compatible.
Push-Button Controls
BATTery: Depress to read battery status.
ON: Power switch. Embedded LED lights when power is applied.
BALance: Activates the auto-balance function. Will set the meter needle to the desired location on the meter face.
SCAN: Activates the automatic lift-off compensation function.
FREQency: Depress to display operating frequency.
Range: Selects between the three sensitivity levels for the unit.

Case Dimensions: (cover closed): 8.4" deep x 11.4" wide x 7" high (21.3cm x 28.9cm x 17.8cm)
Weight: 8.5 lbs. (3.8KG)
Power: Rechargeable Ni-Cad battery. Line operation is user-selectable 115 or 230V.
Enclosure: Case is made of durable aluminum, cover contains accessory packet
Frequency Range: Variable from 55 KHz to 220 KHz
Readout: Rectangular meter, 3.5" wide. Scale numbered from 0 - 500 in 50 divisions
Environment: 0° to 120° at 85% RH